CALCULATION FOR CALIBRATION OF THE DO METER INPUT BP INPUT C O001 LBL O O002 SF 10 O003 DO CALC O004 PSE O005 BP mmHG O006 PSE O007 CF10 O008 INPUT B O009 SF 10 O010 TEMP C O011 PSE O012 CF 10 O013 INPUT T O014 (T + 273.15) O015 STO K O016 EXP (-139.34411+(157570.1/K)-(66423080/K^2)+(12438000000/K^3)-(862194900000/K^4)) O017 STO C O018 EXP(11.8571-(3840.7/K)-(216961/K^2)) O019 STO P O020 (0.000975-(1.426E-005*T)+(6.436E-008*T^2)) O021 STO D O022 (C*(B/760)*(((1-P/(B/760))*(1-D*(B/760)))/((1-P)*(1-D)))) O023 STO N O024 SF 10 O025 CORRECTED DO O026 PSE O027 CF 10 O028 VIEW N O029 SF 10 O030 TEMP C O031 PSE O032 CF 10 O033 VIEW T O034 SF 10 O035 BP MMHG O036 PSE O037 CF 10 O038 VIEW B 0039 RTN K= T + 273.15 C = @EXP(-139.34411+(157570.1/K)-(66423080/K^2)+(12438000000/K^3)-(862194900000/K^4)) PWV = @EXP(11.8571-(3840.7/Keltmp)-(216961/Keltmp^2)) THETA = 0.000975-(1.426E-005*C temp)+(6.436E-008*C temp^2) CP = +C2*($BP/760)*(((1-PWV/($BP/760))*(1-THETA*($BP/760)))/((1-PWV)*(1-THETA)))